Double-End Shear Beam
Description
One of the OEM load cells produced by Camaweigh is a double-end shear beam. Its double-end shear beam configuration makes it insensitive to side load forces. Its recommended applications are low profile platforms, pallet truck, tank, silo weighing, and typical floor scales. This type of load cell is made of nickel-plated alloy steel. Alloy steel is known for its strength and ability to handle large loads.
Built Material
This OEM load cell is made from nickel-plated alloy steel. This ensures the load cell’s improved hardness, durability, toughness, and ability to withstand corrosion.
Accuracy
The weighing accuracy class of this type of Camaweigh’s OEM load cell is rated as OIML C3. The C3 accuracy class rating is given to load cells with high accuracy specification. Load cells in the C3 Accuracy Class classification are mostly used with weighing machines used for tank weighing and light scales.
Maximum Capacity
The maximum force capacity of this OEM load cells ranges from 5t to 50t. Overload is allowed but to a certain extent only.
Safe Overload
The maximum safe overload capacity of this load cell is 150% of its maximum load capacity. Exceeding 150% may put the load cell in danger of being deformed.
Ultimate Overload
The maximum overload allowed is 200% of the maximum load capacity. Exceeding 200% will deform the load cell and cause inaccurate measurements.
Common Problems
Load cells need to be properly mounted in the right position. Improper mounting of the load cells will result into wrong readings and inaccurate results. Overloading weighing device is sometimes unavoidable which may result permanent deformation of the load cell. Load cell permanent deformation will make calibration impossible. Another common problem with load cells is electrical damage. Being exposed to significant electrical currents damage the resistors in the load cells.
Insulation resistance | 50V DC: ≥5000MΩ |
---|---|
Material | nickel-plated alloy steel |
Operating temperature range | -30 to +70℃ |
Combined Error | ±0.03% F.S |
Excitation maximum | 15V DC |
Excitation, recommended voltage | 10 to 12V DC |
Zero balance | ±1% F.S |
Temperature range, compensated | -10 to +40℃ |
Zero Temperature Influence | ±0.02% F.S |
Output Impedance | 702±3Ω; 703±5Ω(cable≥15m) |
Nonlinearity | ±0.02% F.S |
Input Impedance | 750±10Ω |
Cables | 5.2meter(5t);8meter(10t); 10meter(15t); 10/12meter(20t);12/16meter(25t~50t),diameter 6mm |
Environment protection | IP68 |
Ultimate safe overload | 200% OF R.C |
Maximum safe overload | 150% OF R.C |
Creep error(30minutes) | ±0.02% F.S |
Hysteresis error | ±0.02% F.S |
Non-repeatability | ±0.02% F.S |
Rated output | 2.0±0.002mv/v |
Rated capacity | 5ton to 50ton |
* required fields